| Technical Specifications | ||
| Voltage DC | Range | 0 to ±1100 V |
| Best 1 Year, 95% Specification | 3.5 ppm + 2.5 µV | |
| Voltage AC | Range | 22 µV to 1100V |
| 10 Hz to 1 MHz | ||
| Best 1 Year, 95% Specification: | 42 ppm + 8 µV | |
| Resistance | Range | 0 to 100 MΩ, 18 values in x1 and x1.9 |
| Best 1 Year, 95% Specification | 6.5 ppm | |
| Current DC | Range | 0 to ±2.2A (0 to ±11A with 5725A; 0 to ±100 A with 52120A) |
| Best 1 Year, 95% Specification | 35 ppm + 7 nA | |
| Current AC | Range | 9 µA to 2.2A, 10 Hz to 10 kHz (9 µA to ±11A with 5725A; 9 µA to ± 120 A with 52120A) |
| Best 1 Year, 95% Specification | 103 ppm + 8 nA | |
| Wideband ac Voltage option | Range | 300 µV to 3.5V |
| 10 Hz to 50 MHz | ||
| Best 1 Year, 99% Specification | 0.4% + 500 µV | |
| General Specifications | ||
| Warm up time | Twice the time since last warmed up, to a maximum of 30 minutes | |
| Settling time | Less than 5 seconds for all functions and ranges except as noted | |
| Standard interfaces | IEEE-488 (GPIB), RS-232, USB 2.0 Device, Ethernet, 5725A, 52120A, phase lock in (BNC), phase reference out (BNC) | |
| Temperature performance | Operating | 0 °C to 50 °C |
| Calibration:15 °C to 35 °C | 15 °C to 35 °C | |
| Storage | -40 °C to 75 °C | |
| Operating altitude | 2000 m maximum | |
| Relative humidity | Operating | < 80% to 30 °C, < 70% to 40 °C, < 40% to 50 °C |
| Storage | < 95%, non-condensing. | |
| Safety | IEC 61010-1: 300 V CAT II, Pollution Degree 2 | |
| Analog low isolation | 20 V | |
| Electromagnetic environment | IEC 61326-1: Controlled | |
| Line power | 47 Hz to 63 Hz; ±10% 100 V, 110 V, 115 V, 120 V, 200 V, 220 V, 230 V, 240 V | |
| Power consumption | 300 VA | |
| Dimensions | Height | 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet |
| Width | 43.2 cm (17 in), standard rack width | |
| Depth | 64.8 cm (25.5 in), overall; 59.4 cm (23.4 in), rack depth | |
| Weight | 27 kg: (62 lbs.) | |
| Absolute uncertainty definition | 5730A uncertainty specifications include stability, temperature coefficient, linearity, load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated | |
| Specification confidence interval | 99% and 95% | |